ELECTRON MICROSCOPY


Sample

Bovine Complex I

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument HOMEMADE PLUNGER
Cryogen Name ETHANE
Sample Vitrification Details The specimen was vitrified in an environmental plunge-freeze apparatus, blot for 12-15 seconds after the diameter of the blotted meniscus ceases to spread and plunged.
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 19306
Reported Resolution (Å) 5.6
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol RIGID BODY FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details The model from Class1 (D_1200000497) was used for rigid body fitting. The B-factor for individual atoms is carried over from class 1 and doesn't mean anything.
Data Acquisition
Detector Type FEI FALCON II (4k x 4k)
Electron Dose (electrons/Å2) 35
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model FEI TITAN KRIOS
Minimum Defocus (nm) 1800
Maximum Defocus (nm) 5500
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 59000
Calibrated Magnification 105263
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
IMAGE ACQUISITION EPU ?
CTF CORRECTION CTFFIND 4
MODEL FITTING Coot 0.8.3
MODEL REFINEMENT PHENIX 1.10_2155
FINAL EULER ASSIGNMENT RELION 1.4
CLASSIFICATION RELION 1.4
RECONSTRUCTION RELION 1.4
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
NONE Ctf was corrected per particle. Ctf was estimated on the whole micrograph