ELECTRON MICROSCOPY


Sample

ORC-Cdc6 complex of Yarrowia lipolytica bound to a 54bp DNA segment of OriC-061

Specimen Preperation
Sample Aggregation State PARTICLE
Vitrification Instrument LEICA EM GP
Cryogen Name ETHANE
Sample Vitrification Details Sample was applied to a non-glow discharged Quantifoil R 1.2/1.3 300 mesh copper grid (previously washed with ethyl acetate), incubated for 10 seconds, blotted for 2.9 seconds, and plunged into liquid ethane using a Leica Automatic Plunge Freezer EM GP2.
3D Reconstruction
Reconstruction Method SINGLE PARTICLE
Number of Particles 51599
Reported Resolution (Å) 2.73
Resolution Method FSC 0.143 CUT-OFF
Other Details Non-uniform refinement was used for the final reconstruction.
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space REAL
Refinement Protocol FLEXIBLE FIT
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details AlphaFold 2 models for each subunit were docked into the density individually using the "fit to map" functionality in ChimeraX, then refined using Coot. The density for the DNA was sharp enough to allow us to discern purines and pyrimidines, allowing us to produce a generic DNA-B form model of the respective DNA sequence and manually rebuild it in Coot. PHENIX Real Space Refine functionality was used to further process and finalize the structure.
Data Acquisition
Detector Type GATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å2) 43.2
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS KRIOS
Minimum Defocus (nm) 600
Maximum Defocus (nm) 2200
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 105000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details ?
Imaging Experiment
Task Software Package Version
PARTICLE SELECTION Warp ?
MODEL REFINEMENT PHENIX 2.0_5885
RECONSTRUCTION cryoSPARC 4.7
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION CTF correction was done first in WARP during exposure/micrograph pre-processing, and re-corrected during the final refinements/reconstruction of the map in cryoSPARC