ELECTRON MICROSCOPY


Sample

In situ cryo-ET Hibernating Ribosome from S. pombe under 7-day glucose depletion conditions

Specimen Preperation
Sample Aggregation State CELL
Vitrification Instrument LEICA EM GP
Cryogen Name ETHANE
Sample Vitrification Details ?
3D Reconstruction
Reconstruction Method SUBTOMOGRAM AVERAGING
Number of Particles 88206
Reported Resolution (Å) 3.38
Resolution Method FSC 0.143 CUT-OFF
Other Details ?
Refinement Type
Symmetry Type POINT
Map-Model Fitting and Refinement
ID 1
Refinement Space ?
Refinement Protocol ?
Refinement Target ?
Overall B Value ?
Fitting Procedure ?
Details ?
Data Acquisition
Detector Type TFS FALCON 4i (4k x 4k)
Electron Dose (electrons/Å2) 3.4
Imaging Experiment
Date of Experiment ?
Temprature (Kelvin)
Microscope Model TFS KRIOS
Minimum Defocus (nm) 1000
Maximum Defocus (nm) 4000
Minimum Tilt Angle (degrees) ?
Maximum Tilt Angle (degrees) ?
Nominal CS 2.7
Imaging Mode BRIGHT FIELD
Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification 81000
Calibrated Magnification ?
Source FIELD EMISSION GUN
Acceleration Voltage (kV) 300
Imaging Details Objective Aperture 70 um
Imaging Experiment
Task Software Package Version
VOLUME SELECTION Warp 2.0.0dev36
MODEL REFINEMENT PHENIX 1.21.2_5419
CTF CORRECTION Warp 2.0.0dev36
RECONSTRUCTION Warp 2.0.0dev36
Image Processing
CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION 3D CTF Estimation done in Warp
Feedback Form